Reference Intervals for Sysmex XN Hematological Parameters as assessed in the Dutch Lifelines Cohort
Our aim was to derive reference intervals for all Sysmex XN hematology analyzer parameters. The rationale behind the study was the lack of reference intervals for the XN analyzer cell population data (CPD) and functional parameters.
Methods
Fresh fasting blood samples from 18484 participants in the Dutch Lifelines study were analyzed using two automated XN analyzers. Structured health questionnaire data were used to select a subgroup of 15803 apparently healthy individuals for inclusion in the reference population. The method of Latent Abnormal Values Exclusion (LAVE) was used to reduce the influence of latent diseases in the reference population on the resulting reference intervals. We applied analysis of variance to judge the need for partitioning of the reference intervals by sex or age.
Results
We report reference intervals for 105 XN analyzer hematological parameters with and without applying the LAVE method. Sex-related partitioning was required for red blood cells, (RBC, RBC-O), hemoglobin (HGB, HGB-O), hematocrit (HCT), mean corpuscular hemoglobin concentration (MCHC), reticulocyte production index (RPI), and side scattered light intensity of the red blood cell population in the RET channel (RBC-Z). Partitioning for age was not warranted. Body mass index (BMI) and smoking had moderate influence on a minority of the parameters.
Conclusions
We provide reference intervals for all Sysmex XN analyzer routine, CPD and functional parameters, using a direct approach in a large cohort in the Netherlands.